Tests: Long leads modewww.megger.comDLRO2 and DLRO2X37During the test, continuity indicators will fade.The dashes (or an old result) will flash until a new result is displayed.On the secondary display dashes or previous results will flash until a new result is ready.The test will run automatically.If required, to stop the test press the TEST button or break continuity.For a single test the result is displayed constant in the main field. No latest results are displayed in the secondary field unless multiple tests have been made in this sequence.NOTE : The TICK button () will clear the previous results in the secondary fields.Two previous results appear in the secondary fields, left is the previous result and right is the result before that.For auto single tests the continuity indicators are grey, indicating that to repeat the test the leads must be disconnected and reconnected again.If the test has been stopped by the TEST button the continuity indicators will be grey.If continuity already is shown on the C and P connectors press the TEST button to re-start the test.