Tests: Long leads modewww.megger.comDLRO2 and DLRO2X36C and P indicators show red background with for no continuity or green background with for good continuity. 8.1.1 For manual modeProceed when both show green background with . 3. To start the test press the TEST button. During the test, continuity indicators will fade.The dashes (or an old result) will flash until a new result is displayed.On the secondary display dashes or previous results will flash until a new result is ready.For a single test the result is displayed constant in the main field. No latest results are displayed in the secondary field unless multiple tests have been made in this sequence.If continuity is maintained, pressing the TEST button will start a new test.8.1.2 For auto modeIn auto-start mode the instrument will start the test when continuity is established. If both C and P continuity is made before the test mode is selected, the continuity flags will be shown in grey – this means that the user must disconnect and reconnect or press the TEST button to start the test.