DLRO2X mikroohmmeter

Settingswww.megger.comDLRO2 and DLRO2X4111. SettingsIn this section various user settings can be adjusted or accessed. 11.1 General settingsSwitch the rotary switch to the settings mode (Refer to 9.1 General settings on page 28).Go to the general settings tab Megger DLR02 icons μΩμΩmΩ. Use soft keys 1 and 4 to choose the next tab along the top of the settings screen.Use soft keys 2 and 3 to move up and down the items in the general settings tab.General settingsKey Press BeepCan be set on or off.Results BeepCan be set on or off.Background ColourToggle between yellow and white using the tick button.Backlight TimerUse soft keys 2 and 3 to select from off up to 60 seconds (default is 20 seconds).Battery TechnologyToggle between NiMH and alkaline using the tick button.Difference MeterToggle ON and OFF using the tick button.Lower Test LimitChoose μΩ, mΩ or Ω ranging from 1s to 9000s (default is 1 μΩ). Refer to 4.2.5 Setting test limits (DLRO2X) on page 20.Upper Test LimitChoose μΩ, mΩ or Ω ranging from 1s to 9000s (default is 2000 Ω). Refer to 4.2.5 Setting test limits (DLRO2X) on page 20.Pass ColourFor colour blind users the pass colour can be changed from green to blueRestore Factory SettingsSelect using the TICK button () . When the prompt screen appears, press and hold the TICK button () for 3 seconds.CalibrationNot user accessible.11.2 Bluetooth settings (DLRO2X)Go to the bluetooth settings tab .Use soft key 2 to select required setting as listed below.Press the TICK button () to choose which one to edit.Bluetooth settingsBluetoothTurn Bluetooth on or off Bluetooth NameName your unit for connecting to other Bluetooth devices