Tests: Resistance modewww.megger.comDLRO2 and DLRO2X32The dashes (or an old result) will flash until a new result is displayed.During the test, continuity indicators will fade and soft keys will be inactive.For bidirectional tests, the dashes (or the previous result) and left hand arrow will flash while the instrument performs a reading in the forward direction. When the forward measurement is complete, it is shown next to the left hand arrow, and then the right hand arrow and the dashes (or the previous result) flash to indicate that the test current has been reversed.For unidirectional tests, the dashes (or the previous result) will flash until a result is ready. 7.1.1 Post auto-start, unidirectional test:After an auto-start, unidirectional test the C and P continuity indicators will be grey if continuity was maintained. This indicates the C or P lead must be disconnected and reconnected to restart the test.Note: Test button will start another test. Soft key 1 toggles continuous / single test. Soft keys 2 / 3 cycles through current ratings; default is 2 A, selectable 1 A, 100 mA, 10 mA, 1 mA. Soft key 4 is inactive.The TICK button () sets new reference value. 7.1.2 Post auto-start, bidirectional test:After an auto-start, bidirectional test the C and P continuity indicators will be grey. This indicates the C or P lead must be disconnected and reconnected to start the next test.If an auto-start test is interrupted by pressing the TEST button C and P continuity indicator will turn grey. This indicates the C or P lead must be disconnected and reconnected to restart the test. Alternatively press the TEST button to start another test. Continuity on C or P has been lost. Re-establish continuity to start a new test.