DLRO2X mikroohmmeter

Tests: Inductive mode www.megger.comDLRO2 and DLRO2X296.2 Auto stop1. Turn rotary switch to select the inductive mode position. NOTE : To AUTO SAVE results click the SAVE ( ) button now (Refer to 5. Data storage (DLRO2X only) on page 22.).6.2.1 First test Inductive test screen appears, ‘Inductive Mode’ shows in secondary field for a short time then disappears.NOTE : Soft key 1 toggles MANUAL / AUTO. 2. Select AUTO on soft key 1.3. Connect C1-C2 and P1-P2 to the instrument and the unit under test.C and P indicators show red background with for no continuity or green background with for good continuity. Proceed when both show . The continuity indicators are active during the test and will update if continuity is lost.4. To start the test press the TEST button. During the test results will live update on screen. The test will stop automatically when the instrument determines that the reading has been stable for long enough and is unlikely to change significantly; the user can override the instrument and stop the test at any time by pressing the TEST button. The Difference Meter is active. When the test is stopped the displayed result is static.