DLRO2X mikroohmmeter

Safety Warningswww.megger.comDLRO2 and DLRO2X122.5 Safety, Hazard and Warning Symbols on the InstrumentThis paragraph details the various safety and hazard icons on the instrument’s outer case.Icon DescriptionWarning: High Voltage, risk of electric shockCaution: Refer to User Guide.Equipment complies with current EU directives.Equipment complies with current UKCA directives.Equipment complies with current ‘C tick’ requirements.Do not dispose of in the normal waste stream.Equipment protected throughout by double insulation.IP54IP rating2.5.1 Warning IconsThis paragraph details the warning icons that can show on the display.Icon WarningDescriptionExternal Voltage WarningIf an external voltage is applied between the terminals and the instrument is set to ON, the external voltage warning will flash on the display. This is a warning that the item under test is live, it might be dangerous so testing is disabled. The external voltage warning message will flash if more than 5 V potential difference is applied between the voltage terminals and the current terminals. This warning will not show if all terminals are at the same external voltage.NOTE: The warning will not operate if the instrument is set to OFF.Internal Error WarningInternal Error Warning. To clear the error, switch the instrument OFF and then back ON. Contact Megger if this does not clear the error.iRead the User GuideRefer to the user guide if this message shows.